[1]刘立宁.便携式血糖仪与生化分析仪实际方法检测血糖结果比对分析[J].医学信息,2018,31(07):155-156.[doi:10.3969/j.issn.1006-1959.2018.07.054]
 LIU Li-ning.Comparison and Analysis of the Results of Blood Glucose Detection by Portable Blood Glucose Analyzer and Biochemical Analyzer[J].Journal of Medical Information,2018,31(07):155-156.[doi:10.3969/j.issn.1006-1959.2018.07.054]
点击复制

便携式血糖仪与生化分析仪实际方法检测血糖结果比对分析()
分享到:

医学信息[ISSN:1006-1959/CN:61-1278/R]

卷:
31卷
期数:
2018年07期
页码:
155-156
栏目:
诊疗技术
出版日期:
2018-04-01

文章信息/Info

Title:
Comparison and Analysis of the Results of Blood Glucose Detection by Portable Blood Glucose Analyzer and Biochemical Analyzer
文章编号:
1006-1959(2018)07-0155-02
作者:
刘立宁
献县中医医院检验科,河北 献县 062250
Author(s):
LIU Li-ning
Department of Laboratory Medicine,Xianxian Traditional Chinese Medicine Hospital,Xianxian 062250,Hebei,China
关键词:
便携式血糖仪全自动生化分析仪血糖静脉血末梢血准确性
Keywords:
Key words:Portable blood glucose meterAutomatic biochemical analyzerBlood glucoseVenous bloodPeripheral bloodAccuracy
分类号:
R318.6
DOI:
10.3969/j.issn.1006-1959.2018.07.054
文献标志码:
B
摘要:
目的 通过京都GT-1640便携式血糖仪与日立7600全自动生化分析仪测定血糖结果的比对,评价京都GT-1640便携式血糖仪的准确性。方法 选择献县中医医院门诊及住院糖尿病患者100例,同一患者同时采集末梢血和静脉血,分别利用京都GT-1640便携式血糖仪和日立7600全自动生化分析仪检测血糖。结果 京都GT-1640便携式血糖仪与日立7600全自动生化分析仪检测血糖结果,无统计学差异(P>0.05),准确系数>0.981,准确性较好。结论 京都GT-1640便携式血糖仪性能较稳定、重复性较好、准确性较高。
Abstract:
Abstract:Objective The Kyoto GT-1640 portable blood glucose meter and blood glucose results Hitachi 7600 automatic biochemical analyzer alignment accuracy evaluation of Kyoto GT-1640 portable blood glucose meter.Methods The Xianxian county Chinese medicine hospital outpatient and inpatient diabetes patients in 100 cases,the same patients were collected from the peripheral blood and venous blood,respectively using the Kyoto GT-1640 portable blood glucose meter and Hitachi 7600 automatic biochemical analyzer to detect blood glucose.Results Kyoto GT-1640 portable blood glucose meter and Hitachi 7600 automatic biochemical analyzer to detect the blood glucose results,no significant difference(P>0.05),the accuracy coefficient is >0.981 and the accuracy is good.Conclusion Kyoto GT-1640 portable blood glucose meter has stable performance,good repeatability and high accuracy.

参考文献/References:

[1]马琳琳.即时检验血糖仪检测指尖血及静脉全血血糖的临床分析[J].航空航天医学杂志,2016, 27(7):852-853. [2]陈璟.便携式血糖仪与生化仪检测血糖的比较[J].中国医疗前沿杂志,2012,7(10):49-55. [3]中华医学会检验医学分会.便携式血糖仪临床操作和质量管理规范中国专家共识[J].中华医学杂志, 2016,96(36):5. [4]宋国威,李立新,秦英军,等.便携式血糖仪质量管理体系建立的探讨[J].河北医科大学学报,2015, 36(12):1467-1470.

相似文献/References:

[1]丁 星,赵 畅,张海严,等.麦邦数据远程传输型血糖仪临床测量准确性的评价研究[J].医学信息,2018,31(15):64.[doi:10.3969/j.issn.1006-1959.2018.15.020]
 DING Xing,ZHAO Chang,ZHANG Hai-yan,et al.Evaluation of Accuracy of Clinical Measurement of Maibang Remote Data Transmission Glucometer[J].Journal of Medical Information,2018,31(07):64.[doi:10.3969/j.issn.1006-1959.2018.15.020]

更新日期/Last Update: 2018-04-01