[1]丁有得,刘 洋,王 倩,等.自主呼吸状态下光电容积脉搏波法对 健康人群心率变异性的分析[J].医学信息,2019,32(15):163-164.[doi:10.3969/j.issn.1006-1959.2019.15.055]
 DING You-de,LIU Yang,WANG Qian,et al.Analysis of Heart Rate Variability in Healthy Population by Photoelectric Volume Pulse Wave Method under Self-breathing[J].Journal of Medical Information,2019,32(15):163-164.[doi:10.3969/j.issn.1006-1959.2019.15.055]
点击复制

自主呼吸状态下光电容积脉搏波法对 健康人群心率变异性的分析()

医学信息[ISSN:1006-1959/CN:61-1278/R]

卷:
32卷
期数:
2019年15期
页码:
163-164
栏目:
诊疗技术
出版日期:
2019-08-01

文章信息/Info

Title:
Analysis of Heart Rate Variability in Healthy Population by Photoelectric Volume Pulse Wave Method under Self-breathing
文章编号:
1006-1959(2019)15-0163-02
作者:
丁有得刘 洋王 倩佘锦雄巩萍萍
广州医科大学基础医学院生物医学工程系,广东 广州 511436
Author(s):
DING You-deLIU YangWANG QianSHE Jin-xiongGONG Ping-ping
Department of Biomedical Engineering,School of Basic Medical Sciences,Guangzhou Medical University, Guangzhou 511436,Guangdong,China
关键词:
健康人群自主呼吸状态光电容积脉搏波法心率变异性
Keywords:
Key words:Healthy populationSpontaneous breathing statePhotoplethysmography pulse wave method Heart rate variability
分类号:
R318.04
DOI:
10.3969/j.issn.1006-1959.2019.15.055
文献标志码:
A
摘要:
目的 分析自主呼吸状态下光电容积脉搏波法对健康人群心率变异性。方法 选取2016年10月~2018年9月我校本科生健康志愿者64名,所有志愿者均行心电图(ECG)及光电容积脉搏波法(PPG)检测,比较3 min的ECG与PPG检测HRV时的RR间期标准差(SDNN)、RR间期差值均方根值(RMSSD)、低频功率与标准化高频功率的比值(LF/HF),并分析ECG与PPG检测的心率变异性参数的相关性与一致性。结果 64例志愿者3 min的ECG与PPG心率变异性参数比较,差异无统计学意义(P>0.05)。经Pearson相关性分析发现,PPG检测SDNN、RMSSD、LF/HF与ECG各指标呈正相关(SDNN:r=0.999,P<0.001;RMSSD:r=0.998,P<0.001;LF/HF:r=0.998,P<0.001);经可靠性分析发现,PPG用于检测健康人群心率变异性与ECG一致性较高(ICC=0.915,95% CI:0.878~0.943,P<0.001)。结论 健康人群自主呼吸状态下光电容积脉搏波法可用于心率变异性的检测,临床上有较好的应用价值。
Abstract:
Abstract:Objective To analyze the heart rate variability of healthy people under the photoresonance pulse wave method under spontaneous breathing.Methods From October 2016 to September 2018, 64 undergraduate health volunteers were selected.All volunteers underwent electrocardiogram (ECG) and photoplethysmography (PPG), and compared with 3 min ECG and PPG for HRV. RR interval standard deviation (SDNN), RR interval rms value (RMSSD), ratio of low frequency power to normalized high frequency power (LF/HF), and analysis of heart rate variability parameters detected by ECG and PPG Relevance and consistency.Results There were no significant differences in ECG and PPG heart rate variability parameters between the 64 volunteers for 3 min (P>0.05). According to Pearson correlation analysis, PPG detection of SDNN, RMSSD, LF/HF and ECG indicators were positively correlated (SDNN:r=0.999,P<0.001; RMSSD:r=0.998,P<0.001; LF/HF: r=0.998,P<0.001);according to the reliability analysis, PPG was used to detect heart rate variability and ECG consistency in healthy people (ICC=0.915,95%CI:0.878-0.943,P<0.001).Conclusion Photoelectric volume pulse wave method can be used for the detection of heart rate variability in healthy people under spontaneous breathing. It has good clinical application value.

参考文献/References:

[1]石磊,孙朋,庞宇,等.基于光电容积脉搏波描记法的心率变异性分析系统前端装置设计[J].生物医学工程学杂志,2016,33(1):14-17. [2]范哲意,王跃俊,王群.基于光电容积脉搏波的呼吸频率监测[J].北京生物医学工程,2016,35(2):180-184. [3]苌飞霸,陈维平,徐力,等.基于光电容积脉搏波法血氧饱和度测量系统研究[J].工业仪表与自动化装置,2015,45(5):14-16. [4]李浩,吴综,陈真诚,等.基于Android平台的心率变异性检测系统[J].航天医学与医学工程,2017,30(2):98-102. [5]李静,王步青,王卫东.两种计算正常成年人短时心率变异性方法的比较研究[J].北京生物医学工程,2017,36(5):502-506.

相似文献/References:

[1]刘玉民.健康人群的血脂、血糖调查及异常因素分析[J].医学信息,2024,37(24):4.[doi:10.3969/j.issn.1006-1959.2024.24.002]
 LIU Yumin.Investigation of Blood Lipid and Blood Glucose in Healthy People and its Abnormal Factors[J].Journal of Medical Information,2024,37(15):4.[doi:10.3969/j.issn.1006-1959.2024.24.002]

更新日期/Last Update: 2019-08-01