[1]熊武芳,肖光军,刘艳婷,等.GEM premier 3000血气分析仪非传统检测项目与 中心实验室常规仪器检测结果的比较分析[J].医学信息,2018,31(13):160-162.[doi:10.3969/j.issn.1006-1959.2018.13.049]
 XIONG Wu-fang,XIAO Guang-jun,LIU Yan-ting,et al.Comparative Analysis of the Results of Non-traditional Detection of GEM Premier 3000 Blood Gas Analyzer and Conventional Instruments in Central Laboratory[J].Journal of Medical Information,2018,31(13):160-162.[doi:10.3969/j.issn.1006-1959.2018.13.049]
点击复制

GEM premier 3000血气分析仪非传统检测项目与 中心实验室常规仪器检测结果的比较分析()
分享到:

医学信息[ISSN:1006-1959/CN:61-1278/R]

卷:
31卷
期数:
2018年13期
页码:
160-162
栏目:
诊疗技术
出版日期:
2018-07-01

文章信息/Info

Title:
Comparative Analysis of the Results of Non-traditional Detection of GEM Premier 3000 Blood Gas Analyzer and Conventional Instruments in Central Laboratory
文章编号:
1006-1959(2018)13-0160-03
作者:
熊武芳肖光军刘艳婷杨 娜曹 丹陈 姝
遂宁市中心医院检验科,四川 遂宁 629000
Author(s):
XIONG Wu-fangXIAO Guang-junLIU Yan-tingYANG NaCAO DanCHEN Shu
Department of Laboratory,Suining Central Hospital,Suining 629000,Sichuan,China
关键词:
血气分析仪床旁检测中心实验室
Keywords:
Key words:Blood gas analyzerBedside detectionCentral laboratory
分类号:
R446.1
DOI:
10.3969/j.issn.1006-1959.2018.13.049
文献标志码:
A
摘要:
目的 探讨GEM premier 3000血气分析仪动脉血钾、钠、葡萄糖、乳酸、红细胞压积及血红蛋白的检测结果与中心实验室常规仪器Sysmex XN-2000全自动血液分析仪(简称XN-2000)和日立7600全自动生化分析仪(简称日立7600)静脉血检测结果的可比性。方法 以XN-2000和日立7600为参比方法(X),GEM3000为待评方法(Y),分别使用80例患者的静脉血和动脉血进行K+、Na+、Glu、Lac、Hct及Hb的检测,对检测结果进行比对及线性回归分析。结果 GEM3000与N-2000和日立7600相同检测项目的检测结果除Glu和Hb外,K+、Na+、Lac、Hct的差异均具有统计学意义(P<0.05);K+、Na+、Glu、Lac、Hct及Hb的线性回归方程分别为Y=0.9037X+0.2629、Y=0.8868X+11.788、Y=0.9843X+0.0832、Y=0.908X-0.2559、Y=1.0346X-0.3508、Y=1.0095X-0.8035,相关系数(r)分别为0.949、0.875、0.991、0.948、0.917和0.906。结论 GEM3000非传统检测项目仅Glu、Hct和Hb的检测结果与中心实验室常规仪器检测结果一致,能对患者的病情做出快速准确的评估;而K+、Na+和Lac则需建立独立的参考区间,以指导临床更准确的解读其检测结果。
Abstract:
Abstract:Objective To study the results of arterial blood potassium, sodium,glucose and lactic acid,hematocrit and hemoglobin detected by GEM premier 3000 blood gas analyzer,and compared with the results of routine instrument Sysmex XN-2000 automatic Blood Analyzer(XN-2000?for short)and Hitachi 7600 automatic biochemical Analyzer(Hitachi 7600?for short)in central laboratory.Methods Using XN-2000 and Hitachi 7600 as reference method(X)and GEM3000 as the evaluation method(Y),the venous blood and arterial blood of 80 patients were used to detect K+,Na+,Glu,Lac,Hct and Hb respectively.The results were compared and linear regression analysis was made.Results The results of GEM3000 and N-2000 and Hitachi 7600 test items except Glu,Hb,K+,Na+,Lac, Hct were statistically significant(P<0.05);K+,Na+,Glu,Lac,Hct and The linear regression equation of Hb is Y=0.9037X+0.2629,Y=0.8868X+11.788,Y=0.9843X+0.0832,Y=0.908X-0.2559,Y=1.0346X-0.3508,Y=1.0095X-0.8035,respectively.The coefficients(r)are 0.949,0.875,0.991,0.948,0.917,and 0.906,respectively.Conclusion GEM3000 non-traditional testing projects only Glu,Hct and Hb test results are consistent with the results of the central laboratory routine instrumentation,can quickly and accurately assess the patient's condition;and K+,Na+ and Lac need to establish an independent reference interval.To guide the clinical interpretation of the test results more accurately.

参考文献/References:

[1]纪铁梅.血气生化仪与自动生化分析仪在急诊生化检验应用中的比较研究[J].中国医学装备,2016,13(5):103-105. [2]余建洪.GEM Premier 3000血气分析仪非传统检测项目实用价值的探讨[J].检验医学,2016,31(1):26-30. [3]李明辉,时露.动脉血与静脉血中钾离子、钠离子及血糖浓度的比较分析[J].吉林医学,2016,37(10):2424-2425,2426. [4]张秀明.浅析定量检验程序分析性能验证实验方案设计[J].中华检验医学杂志,2015,38(6):428-430. [5]陈剑,周利.室温与4℃下患者动脉血气乳酸水平的比较[J].国际检验医学杂志,2017,38(4):562-563. [6]高阳,张景春,康熙雄.POCT产品在新政策下的质量控制需求[J].中国医疗器械信息,2017,23(11):18-21.

相似文献/References:

[1]刘纲毅,谢妹珍,李丽敏,等.床旁检测心脏型脂肪酸结合蛋白和肌钙蛋白I在急性冠状动脉综合征中的价值[J].医学信息,2021,34(12):93.[doi:10.3969/j.issn.1006-1959.2021.12.024]
 LIU Gang-yi,XIE Mei-zhen,LI Li-min,et al.The Value of Bedside Detection of Cardiac Fatty Acid Binding Protein and Troponin I in Acute Coronary Syndrome[J].Journal of Medical Information,2021,34(13):93.[doi:10.3969/j.issn.1006-1959.2021.12.024]

更新日期/Last Update: 2018-07-01