[1]赵 悦,董 婕,韩晨阳,等.Bioz.com用于心脏手术围术期监测的临床价值研究[J].医学信息,2019,32(05):108-110.[doi:10.3969/j.issn.1006-1959.2019.05.033]
 ZHAO Yue,DONG Jie,HAN Chen-yang,et al.Clinical Value Study of Bioz.com for Perioperative Monitoring of Cardiac Surgery[J].Medical Information,2019,32(05):108-110.[doi:10.3969/j.issn.1006-1959.2019.05.033]
点击复制

Bioz.com用于心脏手术围术期监测的临床价值研究()
分享到:

医学信息[ISSN:1006-1959/CN:61-1278/R]

卷:
32卷
期数:
2019年05期
页码:
108-110
栏目:
临床研究
出版日期:
2019-03-01

文章信息/Info

Title:
Clinical Value Study of Bioz.com for Perioperative Monitoring of Cardiac Surgery
文章编号:
1006-1959(2019)05-0108-03
作者:
赵 悦1董 婕2韩晨阳3周国明1
(1.承德医学院附属医院疼痛科,河北 承德 067000;2.科学出版社,北京 100000;3.承德医学院,河北 承德 067000)
Author(s):
ZHAO Yue1DONG Jie2HAN Chen-yang3ZHOU Guo-ming1
(1.Department of Pain,the Affiliated Hospital of Chengde Medical College,Chengde 067000,Hebei,China;2.Science Press,Beijing 100000,China;3.Chengde Medical College,Chengde 067000,Hebei,China)
关键词:
心脏手术Bioz.com心排血量血流动力学
Keywords:
rdiac surgeryBioz.comCardiac outputHemodynamics
分类号:
R654.2
DOI:
10.3969/j.issn.1006-1959.2019.05.033
文献标志码:
A
摘要:
目的 评价无创血流动力学监测系统(Bioz.com)用于心脏手术患者围手术期心功能评估的准确性及可靠性。方法 选取在2017年12月~2018年6月于承德医学院附属医院行择期心脏手术患者35例,包含瓣膜性疾病23例,冠心病12例。选取“肺动脉漂浮导管”及“超声心动图”作为比较指标,严格按照Bioz.com系统操作规范进行操作,对心排出量(CO)、每搏输出量(SV)、心指数(CI)进行监测,分别记录手术前后患者血流动力学变化情况,并进行相关统计学分析。结果 心脏手术前后,采用 Bioz.com系统及超声心动图两种方法对患者血流动力学参数CO、SV、CI进行监测并比较,差异均无统计学意义(P>0.05)。心脏手术中,采用Bioz.com系统及漂浮导管两种方法对患者血流动力学参数CO、SV、CI进行监测并比较,差异均无统计学意义(P>0.05)。结论 在心脏手术中Bioz.com系统与肺动脉漂浮导管,Bioz.com系统与超声心动图监测的精确度与准确度具有一致性,是一种较为可行的心功能监测方法。
Abstract:
Objectives To evaluate the accuracy and reliability of the non-invasive hemodynamic monitoring system (Bioz.com) for perioperative cardiac function assessment in patients undergoing cardiac surgery. Methods From December 2017 to June 2018, 35 patients underwent elective cardiac surgery at the Affiliated Hospital of Chengde Medical College, including 23 patients with valvular disease and 12 patients with coronary heart disease. The "pulmonary artery floating catheter" and "echocardiogram" were selected as comparison indicators, and the operation was carried out in strict accordance with the Bioz.com system operation specification. The cardiac output (CO), stroke volume (SV), and heart index (CI) were monitored. The hemodynamic changes of patients before and after surgery were recorded and statistical analysis was performed. Results Before and after cardiac surgery, the hemodynamic parameters CO, SV and CI were monitored and compared by Bioz.com system and echocardiography,the differences were not statistically significant (P>0.05). In the cardiac surgery, the hemodynamic parameters CO, SV and CI were monitored and compared by Bioz.com system and floating catheter,the difference was not statistically significant (P>0.05). Conclusion In the cardiac surgery, the Bioz.com system and the pulmonary artery floating catheter, Bioz.com system and echocardiography monitoring are consistent with the accuracy and accuracy, which is a feasible method for heart function monitoring.

参考文献/References:


[1]Caltabeloti F,Monsel A,Arbelot C,et al.Early fluid loading in acute respiratory distress syndrome with septic shock deteriorates lung aeration without impairing arteri al oxygenation: a lung ultrasound observational study[J].Crit Care,2014,18(3):R91.
[2]陈敏芝.超声心动图检查用于心脏瓣膜置换术后的临床价值分析[J].锦州医科大学学报,2017,38(4):25-27.
[3]Urban A,Mace E,Brunner C,et al.Chronic assessment of cerebral hemodyn amics during rat forepaw electrical stimulation using functional ultrasound imaging[J].Neuroimage,2014,101(6):138-149.
[4]Peeters Y,Bernards J,Mekeirele M,et al.Hemodynamic monitoring:To calibrate or not to calibrate? Part1-Calibrated techniques[J].Anaesthesiol Intensive Ther,2015,47(5):487-500.
[5]Bernards J,Mekeirele M,Hoffmann B,et al.Hemodynamic monitoring:To calibrate or not to calibrate? Part2-Non-calibrated techniques[J].Anaesthesiol Intensive Ther,2015,47(5):501-516.
[6]Huang L,Critchley LA.Study to determine the repeatability of supra-sternal Doppler(ultrasound cardiac output monitor) during general anaesthesia: effects of scan quality,flow volume,and increasing age[J].Br J Anaesth,2013,111(6):907-915.
[7]DeMaria AN,Raisinghani A.Comparative overview of cardiacoutput measurement methods: Has impedance cardiographycome of age[J].Congestive Heart Failure,2014,6(2):7-18.
[8]薛华军,马明洲,张铮.超声心排量监测仪与胸阻抗法测定重度心衰患者心功能的比较[J].南京医科大学学报,2013,33(6):849-851.
[9]Squara P,Denjean D,Estagnasie P,et al.Noninvasive cardiac output monitoring(NICOM):a clinical validation[J].Intensive Care Med,2017,33(7):1191-1194.

相似文献/References:

[1]潘闰梅.自制尿管固定装置在心脏手术中的应用[J].医学信息,2018,31(12):184.[doi:10.3969/j.issn.1006-1959.2018.12.060]
 PAN Run-mei.Application of Self-made Urethral Fixation Device in Cardiac Surgery[J].Medical Information,2018,31(05):184.[doi:10.3969/j.issn.1006-1959.2018.12.060]
[2]朱 贺,周国明,韩晨阳.Bioz.com在TURP患者血流动力学变化监测及TURS的早期诊治中的应用[J].医学信息,2019,32(03):74.[doi:10.3969/j.issn.1006-1959.2019.03.023]
 ZHU He,ZHOU Guo-ming,HAN Chen-yang.Application of Bioz.com in Hemodynamic Monitoring of TURP Patients and Early Diagnosis and Treatment of TURS[J].Medical Information,2019,32(05):74.[doi:10.3969/j.issn.1006-1959.2019.03.023]
[3]黄子钊,蒋迎九.心外膜临时起搏在心内直视手术后的应用[J].医学信息,2019,32(10):57.[doi:10.3969/j.issn.1006-1959.2019.10.020]
 HUANG Zi-zhao,Jiang Ying-jiu.Application of Epicardial Temporary Pacing after Open Heart Surgery[J].Medical Information,2019,32(05):57.[doi:10.3969/j.issn.1006-1959.2019.10.020]

更新日期/Last Update: 2019-03-01