[1]沈 骏.自动管电流调制技术对CT血管成像辐射剂量及图像质量的影响[J].医学信息,2026,39(04):140-143,148.[doi:10.3969/j.issn.1006-1959.2026.04.025]
 SHEN Jun.Effect of Automatic Tube Current Modulation Technique on Radiation Doseand Image Quality of CT Angiography[J].Journal of Medical Information,2026,39(04):140-143,148.[doi:10.3969/j.issn.1006-1959.2026.04.025]
点击复制

自动管电流调制技术对CT血管成像辐射剂量及图像质量的影响()

医学信息[ISSN:1006-1959/CN:61-1278/R]

卷:
39卷
期数:
2026年04期
页码:
140-143,148
栏目:
临床证据信息
出版日期:
2026-02-15

文章信息/Info

Title:
Effect of Automatic Tube Current Modulation Technique on Radiation Doseand Image Quality of CT Angiography
文章编号:
1006-1959(2026)04-0140-05
作者:
沈 骏
泰州市疾病预防控制中心职业健康科,江苏 泰州 225300
Author(s):
SHEN Jun
Occupational Health Department, Taizhou Center for Disease Control and Prevention, Taizhou 225300, Jiangsu, China
关键词:
自动管电流调制技术CT血管成像辐射剂量图像质量病变检出率
Keywords:
Automatic tube current modulation technology CT angiography Radiation dose Image quality Lesion detection rate
分类号:
R445
DOI:
10.3969/j.issn.1006-1959.2026.04.025
文献标志码:
A
摘要:
目的 探讨自动管电流调制技术在CT血管成像(CTA)检查中对辐射剂量、图像质量、病变检出率及患者耐受度的影响。方法 选取2022年10月-2024年10月泰州市某三甲医院收治的280例需行CTA检查的患者作为研究对象,采用随机数字表法分为对照组和观察组,每组140例。对照组采用固定管电流模式进行CTA扫描,观察组采用自动管电流调制技术进行CTA扫描。比较两组辐射剂量相关指标(容积CT剂量指数、剂量长度乘积、有效剂量)、图像质量相关指标(图像质量评分、CT值、噪声、信噪比、对比噪声比、图像质量达标率)、病变检出率及患者耐受度指标(不良反应发生率、视觉模拟评分)。结果 观察组容积CT剂量指数、剂量长度乘积、有效剂量均低于对照组(P<0.05)。观察组图像质量评分、信噪比、对比噪声比均高于对照组(P<0.05),且图像质量达标率(98.57%)与对照组(96.43%)相比,差异无统计学意义(P>0.05)。观察组病变检出率(98.86%)高于对照组(91.86%)(P<0.05)。观察组视觉模拟评分低于对照组(P<0.05),不良反应发生率(4.29%)与对照组(5.71%)相比,差异无统计学意义(P>0.05)。结论 自动管电流调制技术应用于CTA检查可降低辐射剂量,提升病变检出率和患者耐受度,同时维持良好的图像质量,符合临床低剂量高质量的检查需求,具有重要的临床推广价值。
Abstract:
Objective To investigate the effect of automatic tube current modulation technique on radiation dose, image quality, lesion detection rate and patient tolerance in CT angiography (CTA). Methods A total of 280 patients who needed CTA examination in a tertiary hospital in Taizhou from October 2022 to October 2024 were selected as the research objects. They were divided into control group and observation group by random number table method, with 140 patients in each group. The control group was scanned by CTA with fixed tube current mode, and the observation group was scanned by CTA with automatic tube current modulation technology. The radiation dose related indexes (volume CT dose index, dose length product, effective dose), image quality related indexes (image quality score, CT value, noise, signal-to-noise ratio, contrast-to-noise ratio, image quality compliance rate), lesion detection rate and patient tolerance index (incidence of adverse reactions, visual analogue score) were compared between the two groups. Results The volume CT dose index, dose length product and effective dose of the observation group were lower than those of the control group (P<0.05). The image quality score, signal-to-noise ratio and contrast-to-noise ratio of the observation group were higher than those of the control group (P<0.05), and the image quality compliance rate (98.57%) was not significantly different from that of the control group (96.43%) (P>0.05). The detection rate of lesions in the observation group (98.86%) was higher than that in the control group (91.86%) (P<0.05). The Visual Analogue Scale score of the observation group was lower than that of the control group (P<0.05). There was no significant difference in the incidence of adverse reactions between the observation group (4.29%) and the control group (5.71%) (P>0.05). Conclusion The application of automatic tube current modulation technology in CTA examination can reduce the radiation dose, improve the lesion detection rate and patient tolerance, while maintaining good image quality, which meets the clinical needs of low-dose and high-quality examination, and has important clinical promotion value.

参考文献/References:

[1]黄涛,马隆佰,陆玉敏,等.80kV管电压下迭代重建算法联合自动管电流调制技术在低剂量头颈部CT血管造影中的应用[J].广西医学,2020,42(18):2364-2368.[2]李冠,邹明宇,王金宝,等.kVassist技术联合自动管电流调制技术及ASIR重组算法在头颈CTA成像的可行性研究[J].临床放射学杂志,2018,37(5):865-868.[3]付小娇,刘芳,高超,等.320-DVCT双低技术在肺动脉栓塞诊断中的可行性研究[J].医学研究杂志,2018,47(10):145-149,153.[4]靖婧,苗延巍.优化低剂量一站式全脑CT灌注成像联合CTA扫描方案[J].中国医学影像技术,2018,34(3):424-428.[5]王振宇,刘琼,赵田田,等.基于ATCM技术的不同探测器宽度对头颈CT血管成像图像质量与辐射剂量影响的研究[J].中国医疗设备,2024,39(6):131-136.[6]葛怀志,吴爱琴,陈久尊,等.低碘浓度对比剂结合Z-dom及iDose4技术在颈部CT血管造影中的应用[J].温州医科大学学报,2020,50(11):901-905.[7]刘彬,高思喆,王传卓,等.低管电压联合全模型迭代重建技术在下肢动脉CTA中的应用[J].中国临床医学影像杂志,2018,29(9):660-663.[8]张晓龙,赵英明,张国祥,等.头颈部CTA中GSI扫描联合“双低”技术在中老年人群体的应用价值[J].CT理论与应用研究,2025,34(2):295-302.[9]牛丹丹,李莹,张颖颖,等.Auto-kV技术联合ClearView+迭代算法在主动脉CTA的可行性研究[J].临床放射学杂志,2020,39(5):999-1003.[10]原媛,钟朝辉,王振常,等.器官剂量调制技术在头颈CTA中对辐射敏感器官的照射强度和图像质量的影响[J].临床和实验医学杂志,2019,18(6):645-649.[11]陈岩,王建华,高希法,等.基于全模型迭代重建算法的低剂量胸痛三联CTA的应用[J].生物医学工程与临床,2022,26(3):326-332.[12]袁颖,白玫,乔磊,等.基于自动管电流调制技术偏中心定位对胸部CT图像质量和辐射剂量的影响[J].中国医学装备,2023,20(12):5-9.[13]袁颖,吴天棋,钟朝辉.应用自动管电流调制技术不同探测器宽度对胸部CT图像质量及辐射剂量的影响[J].中国医学装备,2019,16(11):19-22.[14]王菲菲,罗娅红.自动管电流调制低剂量CT技术检查腮腺肿瘤的临床价值[J].医学影像学杂志,2019,29(10):1677-1681.[15]王静,梁保辉,王梦廷,等.定位像参数对智能管电压辅助技术联合自动管电流调制技术性能的影响[J].中国医疗设备,2023,38(3):21-25.[16]李翔,王翔,杨欣,等.自动管电流技术联合人工智能在COVID-19肺部低剂量CT中的应用价值[J].放射学实践,2020,35(9):1082-1086.[17]王文生,贾卫伟,王立艳,等.kVAssist2.0联合自动管电流调制在256排CT腹盆部检查中的应用研究[J].中国中西医结合影像学杂志,2024,22(5):611-614,621.[18]张曼曼,王静,王梦廷,等.摆位偏差对管电流调制CT检查辐射剂量和图像质量的影响[J].山西医药杂志,2022,51(18):2047-2051.[19]张鹏,都基权,许韫,等.腹部CT低剂量扫描自动管电流调制技术的应用价值[J].武警医学,2020,31(4):287-290.[20]朱晨晨,马梓轩,冯泽臣,等.定位像球管投照角度与不同管电压组合对头部CT辐射剂量与图像质量的影响:基于ATCM联合能谱纯化技术[J].CT理论与应用研究,2025,34(5):884-890.

更新日期/Last Update: 1900-01-01